Large Aliquot Injection Approach in the Electrothermal AAS Determination of Arsenic in Environmental Water Samples |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
著者 |
IMAI, Shoji
FUJIKAWA, Kenichiro
YONETANI, Akira
OGAWA, Nobuaki
KIKUCHI, Yoichi
|
著者(機関) |
|
|
値 |
Department of Chemistry, Faculty of Integrated Arts and Sciences, The University of Tokushima |
著者(機関) |
|
|
値 |
Department of Chemistry, Faculty of Integrated Arts and Sciences, The University of Tokushima |
著者(機関) |
|
|
値 |
Hitachi Instruments Engineering Co. Ltd. |
著者(機関) |
|
|
値 |
Department of Materials-process Engineering & Applied Chemistry for Environment, Faculty of Engineering and Resource Science, Akita University |
著者(機関) |
|
|
値 |
Department of Chemistry, Faculty of Education, Iwate University |
登録日 |
|
|
日付 |
2009-12-16 |
書誌情報 |
Analytical Sciences
巻 16,
号 2,
p. 163-167,
発行日 2000-01-01
|
ISSN |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
0910-6340 |
Abstract |
|
|
内容記述タイプ |
Other |
|
内容記述 |
Determination of As by direct electrothermal atomization atomic absorption spectrometry with a pyropytically-coated graphite furnace treated with W by a single drop treatment technique was reported. The treated PG furnace made it possible to introduce a 100 µl aliquot of sample solution into the furnace with a high precision (RSD <4%). Use of a Pd-modifier made it possible to enhance the precision (RSD <3%) and the sensitivity. A good linearity (correlation factor >0.99) between integrated absorbance and injection volume was observed in the range 0 - 100 µl. The calculated detection limit and characteristic mass (sensitivity) for integrated absorbance were 22 ng l^-1 and 16 pg and for maximum absorbance were 45 ng l^-1 and 13 pg for a 100 µl injection with the Pd modifier at the optimum ashing temperature of 800°C. An interference study for environmental matrices was also carried out. The recoveries of 2 and/or 4 µg l^-1 of As added to various water samples, such as snow, cloud, river and tap waters, were 97 - 104%. |
出版者 |
|
|
出版者 |
The Japan Society for Analytical Chemistry |
権利 |
|
|
権利情報 |
Copyright (c) 2000 The Japan Society for Analytical Chemistry |
DOI |
|
|
関連タイプ |
isIdenticalTo |
|
|
識別子タイプ |
DOI |
|
|
関連識別子 |
10.2116/analsci.16.163 |
著者版フラグ |
|
|
出版タイプ |
VoR |
|
出版タイプResource |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |